Technical parameters
Product Size 153mm (L) × 162mm (W) × 227mm (H) Allowable error ± 5%
Product Weight 1.7kg
Mode Selection drip rate mode, volume mode, time mode
Infusion Set Specification special or standard disposable infusion set
History recording function, maximum 2000 history can be stored.  
Infusion Rate Range ml/h Mode: (1 to 1200) ml/h, 1ml/h increment per step Drop/min Mode: (1 to 266)  drops/min
Infusion Rate Accuracy ml/h mode: within ±5% (after calibration)
Drop/min Mode within ±3% (after calibration)
Mechanical Accuracy of Infusion Pumps within ±2%
Rapid Infusion Rate 100 ml/h~1000 ml/h (100 ml/h per step increment)
KVO Speed 1ml/h~5ml/h (each increment is 1ml/h)
Infusion volume 1ml~9999ml
Cumulative Infusion Volume  0ml~9999.9ml
Blocking alarm threshold Height: 800mmHg±200mmHg (106.7kPa±26.7kPa)
  Medium: 500mmHg±100mmHg (66.7kPa±13.3kPa)
  Low: 300mmHg±100mmHg (40.0kPa±13.3kPa)
Alarm Infusion completed, pipeline blocked, door opened, air bubbles, battery low, etc.
Power Supply 220VAC , 50Hz; built-in battery: 11.1V rechargeable lithium battery pack, capacity ≥1600mAh; after the new battery is fully charged, the infusion pump can work continuously for more than 4 hours at a rate of 25ml/h.
Bubble Detector ultrasonic detection mode, detection sensitivity ≥25μL.
Fuse specification T2AL/250VAC, quantity 2, installed in infusion pumps
IP Rating IPX2 splash-proof equipment
Safety Class Class I, internal power supply, CF type continuous operation equipment
Operating environment conditions: a) Ambient temperature range +5℃~+40℃
b)Relative Humidity 20~90%
c) Atmospheric Pressure Range 86.0KPa~106.0 KPa
Transportation and storage conditions: a) Ambient temperature range -20 ℃ ~ +55 ℃
b) Relative humidity  ≤93%